Research Article: Consistency and correlation of a domestic Kehua serum free light chain assay with two imported analytical systems
Abstract:
Serum free light chain (sFLC) testing is central to the diagnosis and monitoring of plasma cell disorders, but assay availability and implementation can vary by laboratory setting. We evaluated a domestic latex-enhanced immunoturbidimetric Kehua serum sFLC assay against Freelite and N Latex systems.
Residual serum samples from 133 adults with suspected or established monoclonal gammopathies were measured in parallel by Kehua and Freelite; 122 were also tested by N Latex. Precision was assessed with low- and high-level serum controls. Method agreement was evaluated using correlation, Passing-Bablok regression, Bland-Altman analysis, qualitative concordance and comparator-defined ROC analysis.
Kehua within-run CVs were <5% for ? and ? controls. Agreement with Freelite was excellent for FLC-?, FLC-? and ?/? ratio (Pearson r = 0.988-1.000; CCC = 0.983-0.999), with Passing-Bablok slopes close to unity. Agreement with N Latex was also strong (r = 0.991-0.998), although ?/? ratio showed proportional bias. Total qualitative concordance was 89.47-93.23% versus Freelite and 83.61-91.80% versus N Latex, with Cohen’s ? of 0.67-0.85. Comparator-defined ROC AUCs were 0.994-0.999 for elevated FLC-? or FLC-? and 0.796-0.869 for ?/? ratio abnormality.
The Kehua serum assay demonstrated acceptable precision and good clinical agreement with two imported methods. It may serve as a domestic option for routine serum sFLC quantification, provided method-specific intervals are maintained and the same platform is used for serial monitoring.
Introduction:
Serum free light chain (sFLC) testing is central to the diagnosis and monitoring of plasma cell disorders, but assay availability and implementation can vary by laboratory setting. We evaluated a domestic latex-enhanced immunoturbidimetric Kehua serum sFLC assay against Freelite and N Latex systems.